IMPACT 754 VENTILATOR:
PRE-MISSION CHECKS AND TROUBLESHOOTING
11. Turn the HIGH Pressure alarm to 100.
12. Set LOW pressure alarm to 5cmH O above PEAK.
2
a. The ventilator will display the following and trigger an alarm.
“LOW PRESSURE PEAK INSPIRATORY PRESSURE TOO LOW”
13. Remove the circuit from the ventilator.
a. The ventilator will display “DISCONNECT CHECK CIRCUIT CONNECTIONS” and
trigger an alarm.
14. Preset ventilator for use per Duty Inspection, Power On Checks, and turn-off.
190 • PROCEDURES