Page 198 - 2026 SMOG
P. 198

IMPACT 754 VENTILATOR:
                PRE-MISSION CHECKS AND TROUBLESHOOTING
         11.  Turn the HIGH Pressure alarm to 100.
         12.  Set LOW pressure alarm to 5cmH O above PEAK.
                                  2
            a.  The ventilator will display the following and trigger an alarm.
              “LOW PRESSURE PEAK INSPIRATORY PRESSURE TOO LOW”











         13.  Remove the circuit from the ventilator.
            a.  The ventilator will display “DISCONNECT CHECK CIRCUIT CONNECTIONS” and
              trigger an alarm.













         14.  Preset ventilator for use per Duty Inspection, Power On Checks, and turn-off.















        190  •   PROCEDURES
   193   194   195   196   197   198   199   200   201   202   203