Page 195 - 2026 SMOG
P. 195

IMPACT 754 VENTILATOR:
                PRE-MISSION CHECKS AND TROUBLESHOOTING
         3. Check BATT OK














         4. Preset ventilator knobs to:
           a.  Rate (2) 18
           b.  Inspiration Time (3) 1:2
           c.  Vt (4) 500
           d. FiO  (5) 100%
               2
           e.  High-pressure alarm to 35cmH O
                                 2
           f.  LOW pressure to 15cmH O
                             2







         5. Turn OFF
         6. Store Ventilator with Air Inlet and Gas Out Ports protected and covered.











           Ventilator is now pre set for duty and able to be rapidly employed as needed with minor
           adjustments to Vt based on patient’s ideal body weight and turning on O  source.
                                                         2
                                                                   (continues)