IMPACT 754 VENTILATOR:
PRE-MISSION CHECKS AND TROUBLESHOOTING
Routine Care
Clean the unit and hose attachments with a damp, soapy cloth and wipe dry. The inlet filter may be
removed to check for dirt or debris. Check metal hose couplings for thread wear and debris.
Duty Inspection
Power Off Checks
1. Within the calibration date (6-month maintenance cycle)
2. Air inlet clear and filter in place (Right side of vent)
3. Gas (“OXYGEN IN” and “AIR IN”) and Patient (“EXHALATION VALVE” and
“TRANSDUCER”) connections are clear and tight (Top of vent)
(continues)