IMPACT 754 VENTILATOR:
                PRE-MISSION CHECKS AND TROUBLESHOOTING
                                  Power On Checks
         1. Turn the "MODE" knob (1) to the desired setting (A/C, SIMV, CPAP)














         2. The ventilator will run a SELF-TEST upon setup. CAUTION: SELF  CHECK must be
           performed with the disposable ventilator circuit disconnected.
           **Ignoring this requirement could cause the SELF-CHECK process to sense a residual
           airway pressure, leading to a SELF CHECK failure.**
           At this point, CAL is not required. If SELF-TEST results in a Calibration Failure, place (1) to
           CAL until CAL OK is displayed. If calibration fails, the ventilator is deadlined.