IMPACT 754 VENTILATOR:
                PRE-MISSION CHECKS AND TROUBLESHOOTING
                                 Weekly Inspections
         **ALL CAUTIONS, WARNINGS, AND NOTIFICATIONS THAT CORRESPOND WITH
         THE 754 SCREEN WILL BE IN ALL CAPS AND BOLD**
         1. Complete Duty Inspection
         2. Set FiO  to 21%
                2
         3. Attach the vent circuit to the vent and field expedient training lung:
           a.  Slide one large exam glove inside another.
           b.  Wrap the open end of the gloves around the patient end of the circuit tube.
           c.  Secure with rubber band (DD1380)/Cut off wrist bead of another glove/Penrose drain/Tape.












         4. Connect the high-pressure oxygen source and turn on O 2
         5. Turn the vent on and allow respirations to begin (listen for the compressor)
         6. Set FiO  to 100%
                2
           a.  Internal compressor will stop (audible)
         2. Turn off the oxygen source
           a.  The ventilator will alarm and show
                  “O  LOW/FAIL CHECK OXYGEN SOURCE/CONNECTIONS”
                    2
                                       and
                                “FiO -GAS MIX ERROR.
                                   2
                       CHECK SOURCE/SETTINGS/CONNECTIONS”
           b.  Compressor will turn on (audible)












                                                                   (continues)