IMPACT 754 VENTILATOR:
                PRE-MISSION CHECKS AND TROUBLESHOOTING
          8.  Set FiO  to 21%.
                 2
          9.  Set HIGH pressure alarm to 5cmH O below PEAK.
                                  2
            a.  The ventilator will “stutter” and show a “HIGH PRESSURE PEAK INSPIRATORY
              PRESSURE TOO HIGH” warning and signal an alarm.












         10.  Turn PLATEAU pressure ON.
            a.  Ventilator will display the following and trigger an alarm:
                         “HIGH PRESSURE PEAK INSPIRATORY.
                                PRESSURE TOO HIGH”
                            “PLATEAU VOLUME DELIVERED
                           VOLUME LESS THAN SET VOLUME”


















                                                                   (continues)






                          2026 Standard Medical Operating Guidelines (SMOG)  •   189