Page 13 - JSOM Fall 2022
P. 13

FIGURE 5  Post-drop.                               Results
                                                                 Each SATB parachute deployed appropriately for all 12 air­
                                                                 drops. There was minor damage to the stitching on the blood
                                                                 cooler strap system on half of the airdropped coolers, but
                                                                 other wise, there was no visible damage to the blood coolers or
                                                                 cooling elements. All WB units within the cooler were intact
                                                                 except for one unit on the C­145 drop that had minor leakage.
                                                                 The WB unit with minor damage was discarded without post­
                                                                 drop laboratory testing. Neither the experimental nor control
                                                                 WB units showed any evidence of hemolysis on gross exam­
                                                                 ination; neither were schistocytes noted on peripheral smear.

                                                                 Accelerometer data were collected for each of the airdropped
                                                                 units and compared with the data for the control (Table 1).
              (A) Retrieving bundles after the C­145 drop. (B) Post­drop laboratory   The temperature within the test and control coolers remained
              testing. (C) Post­drop laboratory testing. (D) Minor damage to a box.   between 6.82°C and 8.35°C, and the average pressure was be­
              (E) Intact WB units post­drop. (F) Bundle on drop zone. (G) Retrieving
              bundles after the C­130 drop.                      tween 997.5 mbar to 1012.0 mbar. The average mechanical
                                                                 shock for the C­145 airdrop recorded 377.6 Gz, compared
                                                                 with the C­130 airdrop, which experienced 197.2 Gz. The av­
              were added to the front pocket, as seen in Figure 1E. For the
              C­130 drops, paratroop door delivery was selected because   erage shock exerted on the control group was 36.8 Gz.
              of the lower 11 psf requirement for door bundles. No added
              weight was required. For each aircraft, an exception issued by   TABLE 1  G-Meter Data
              the Fort Lee Aerial Delivery  and Field  Services Department      Average      Average   Average Total
              lowered the minimum required weight for the chute from    Factor  Temperature (°C) Pressure (mbar)  Shock (g)
              30 lb to 13.5 lb.                                  Control       8.35 ± 0.16  1012.0 ± 0.8  36.8 ± 9.1
                                                                 C­145 airdrop  6.62 ± 1.12  997.5 ± 4.8  377.6 ± 82.6
              The blood coolers were airdropped by the C­145 at an altitude   C­130 airdrop  8.16 ± 0.99  998.8 ± 7.0  197.2 ± 57.3
              of 200 feet above ground level (AGL) flying at approximately
              100 nautical miles per hour. For the C­130, the blood coolers   Mean baseline and post­airdrop laboratory analyses are sum­
              were airdropped from the paratroop door at an altitude of   marized in Table 2 and Figure 6. Notably, there was a 2× dif­
              500 feet AGL, flying at approximately 130 nautical miles per   ference in potassium measurements between the C­145 airdrop
              hour. Bundles were immediately collected from the drop zone   and both the control and C­130 airdrops. However, there was
              and ground­transported to the Eglin Hospital Laboratory for   a consistent increase between baseline and post­intervention
              post­airdrop laboratory analysis and storage.      potassium measurements in each airdrop and control group.
                                                                 Additionally, the hemoglobin levels between the pre­drop and
              Storage, Sampling, and Testing                     post­drop laboratory values for the C­130 indicated an aver­
              All WB units were stored at 3°C. Each WB unit was individu­  age decrease from 13.71g/dL to 11.50g/dL, but the C­145 air­
              ally mixed by gentle inversion prior to sampling. Open system   drop and controls seemed to have small increases of 11.76g/dL
              sampling of 16mL WB, aliquoted into four 13 × 75mm plas­  to 11.83g/dL and 12.58g/dL to 12.78g/dL, pre­ to post­
              tic tubes, was conducted before and after the airdrop. One   intervention, respectively. Similarly, the respective hematocrit
              tube was delivered to the coagulation, hematology, chemistry,   baseline and post­airdrop levels were 41.32% and 35.25% for
              and blood gas sections for respective laboratory analysis. Co­  the C­130 drop, 36.3% and 37.24% for the C­145 drop, and
              agulation studies were performed on the STA Compact Max   38.2% and 38.96% for the control group. Furthermore, the
              analyzer (Stago). Complete blood counts were performed on   platelet counts increased in each airdrop test group between
              the DxH 800 Hematology Analyzer (Beckman Coulter). The   the baseline and post­intervention laboratory values. Platelet
                                                                                                  –3
              GEM Premier 4000 analyzer (Instrumentation Laboratory)   counts for the C­130 drop were 96.46 10 /µL pre­drop and
              was used to measure WB pH. Lactate, potassium, LDH, bili­  130.33 10 /µL post­drop; C­145 counts were 58.7 10 /µL
                                                                                                             –3
                                                                         –3
                                                                                 –3
              rubin, and glucose measurements were performed on the At­  pre­drop and 67.3 10 /µL post­drop; and control counts were
                                                                        –3
                                                                                              –3
              telica integrated chemistry analyzer (Siemens). WB units were   88.47 10 /µL pre­drop and 99.47 10 /µL post­intervention.
              packaged in Collins boxes for transportation to the airfield.   Lactate increased between baseline and post­intervention lab­
              Post­drop, units were received in a BloodBoxx container. The   oratory values for each airdrop and the control.
              temperature of the experimental and control units was 7.6°C
              and 6.8°C, respectively. Post­sampling was accomplished   An ANCOVA test was performed to compare post­interven­
              within 24 hours of baseline sampling.              tion results between the C­130 airdrop, C­145 airdrop, and
                                                                 no­ airdrop control blood units by controlling for the baseline.
              Statistical Analysis                               Tukey’s post hoc test was performed to learn whether a statis­
              Analysis of covariance (ANCOVA) was used to compare the   tical significance existed between different means. This analy­
              group differences, controlling for baseline. Tukey’s post hoc   sis revealed that there were statistically significant differences
              tests were used to adjust for the number of tests performed to   in the means for potassium levels between the C­145 airdrop
              control for type I errors, and we reported adjusted p values.   and control, as well as the C­145 and the C­130 airdrop. Also,
              The p value <.05 was considered statistically significant. Sta­  there was a significant difference between the total bilirubin
              tistical analyses were performed using SAS version 9.4 (Statis­  and indirect bilirubin in the C­130 airdrop versus the C­145
              tical Analysis Software).                          airdrop. LDH and PTT in the C­130 versus the C­145 airdrop,

                                                                        Effect of Airdrop on Fresh and Stored Whole Blood  |  11
   8   9   10   11   12   13   14   15   16   17   18