Page 132 - JSOM Fall 2018
P. 132

FIGURE 12  Patient one-way nonrebreather valves, the Ambu E   result in negative pressure in the draw­over apparatus when
          valve on the left and the Laerdal valve on the right.  the oxygen flow is less than the patient’s minute volume, lead­
                                                             ing to a collapse of the SIB and loss of ventilation. A blocked
                                                             reservoir with an oxygen flow greater than the patient’s min­
                                                             ute volume will result in pressurization of the draw­over va­
                                                             porizer, leading it to act like a plenum vaporizer. The end of
                                                             the reservoir should also be secured to avoid it dropping onto
                                                             the floor where dust and contamination may be drawn up into
                                                             the system and delivered to the patient’s respiratory system.


                                                             Mechanical Ventilation
                                                             In the very austere setting, ventilation can be performed solely
                                                             using an SIB. A nonskilled assistant can be readily instructed
                                                             in the use of the SIB in order to free the anesthetist for more
                                                             important tasks. However, the manual ventilation of a patient
                                                             can become quite tiring over the duration of an anesthetic, and
                                                             it is unlikely to produce as consistent a minute volume as a
          oxygenation. A length of anesthetic tubing can also be attached   mechanical ventilator. This is of importance in the delivery of
          to the expiratory end of the PEEP valve to vent expired gasses   protective lung ventilation, which is of particular significance
          away from the surgical team, thus acting as a primitive scav­  in the treatment of patients with pulmonary injury resulting
          enging system. This can be important if these devices are used   from blast or inhalational injuries. 29
          for prolonged periods in poorly ventilated spaces to reduce the
          surgical team’s exposure to expired anesthetic vapor.  A mechanical ventilator can be used with a draw­over vapor­
                                                             izer in either the draw­over or push­over configurations.  In
                                                                                                          30
                                                             the draw­over configuration, it is positioned between the va­
          Oxygen Reservoir
                                                             porizer and the patient, and in the more common push­over
          Volatile anesthetic agents have a number of effects on ven­  configuration, the ventilator is attached to the vaporizer in­
          tilation. They decrease tidal volume, depress the response to   let in the place of the oxygen reservoir. 31,32  Both the PAC and
          hypercarbia, inhibit hypoxic pulmonary vasoconstriction, and   OMV vaporizers can have a ventilator operated in both po­
          increase ventilation perfusion mismatching and pulmonary   sitions with minimal changes in the volatile output.  When
                                                                                                      31
                                                         24
          shunt. All of these effects can decrease oxygen saturation.    using a ventilator in push­over mode, the SIB should be re­
          This is particularly relevant in patients with preexisting pul­  moved from the circuit in order to reduce the compliance of
          monary disease or traumatic chest injury. When available,   the system. The benefit of using the ventilator in the push­over
          oxygen should be added to the carrier gas delivered to the va­  position is that the internal mechanics and seals of the ventila­
          porizer. Adding it upstream of the vaporizer will avoid diluting   tor do not get exposed to volatile anesthetic, which can cause
          the anesthetic delivered to the patient.           malfunction of some ventilators. 32
          Spontaneous and mechanical ventilation can be divided into   Monitoring the Patient
          the four phases of inspiration, expiration, and inspiratory and
          expiratory pauses. During inspiration, the peak inspiratory   The use of capnography monitoring has vastly improved the
          flow developed by an adult male is 300 to 500L/min,  well in   safety profile of anesthesia  and the development of porta­
                                                   25
                                                                                  33
          excess of the flow available from an oxygen tank or concentra­  ble battery operated capnographs  means this technology is
                                                                                        34
          tor. This is why it is impossible to deliver more than 35% to   available for use with draw­over anesthesia when access to
          40% Fio  using a Hudson face mask even with an oxygen flow   power is not guaranteed. Portable mainstream capnographs
                 2
          rate of 15L/min. Using a draw­over without a reservoir, air will   are available that can clip onto the end of an endotracheal
          be drawn into the circuit, diluting down the added oxygen. An   tube and operate on a single AAA battery. Use of capnography
          oxygen reservoir placed upstream of the oxygen inlet will catch   reduces the risk of under or over ventilating a patient and,
          the oxygen lost to the environment during the expiratory phase   in addition, monitoring capnography adds the reassurance of
          of respiration, allowing it to be delivered to the patient during   apnea alarms warning of failed intubation, patient apnea, and
          inspiration. A  reservoir  can  be made from  the reservoir bag   disconnection.
          that comes with the Laerdal and Ambu SIB or from a length
          of anesthetic “elephant” tubing. One meter of this tubing has   End­tidal agent monitoring is also available in lightweight
          an internal volume of approximately 500mL. With an oxygen   battery­operated devices. The multigas monitor by Acutronic
          flow rate of 4L/min, a 500mL oxygen reservoir will deliver an   gives graphic representation of volatile anesthetic concen­
          Fio to the patient of approximately 70% to 80%. At a flow   tration and CO  along with a pulse oximeter. The ability to
            2
                                                                         2
          of 1L/min, the resulting Fio is approximately 30%.  A C­size   monitor end­tidal anesthetic concentration allows the deliv­
                                                  26
                               2
          oxygen cylinder will last up to 400 minutes with a flow of 1L/  ery of a precise level of anesthetic as opposed to estimating
          min.  Oxygen can be supplied to the circuit using either bot­  the delivered concentration based on vaporizer dial settings,
             27
          tled oxygen or an oxygen concentrator. Most oxygen concen­  temperature,  and  the patient’s  minute  volume. This  reduces
          trators are able to deliver 5 to 6L of 95% oxygen. 28  the risk of delivery of too deep an anesthetic and the risk of
                                                             awareness from inadequate volatile agent. Monitoring the
          When using anesthetic tubing as a reservoir, it is important   end­tidal agent concentration should also provide a warning
          to ensure that the end of the tube is not blocked as this may   of an empty vaporizer, a significant benefit given the small
          130  |  JSOM   Volume 18, Edition 3 / Fall 2018
   127   128   129   130   131   132   133   134   135   136   137