11.Turn HIGH Pressure alarm to 100.
          12.Set LOW pressure alarm to 5cmH 2 O above PEAK
               a. Ventilator will display “LOW PRESSURE-PEAK INSPIRATORY PRESSURE TOO LOW” and
                 trigger an alarm.


















          13. Remove circuit from ventilator.

               a. Ventilator will display “DISCONNECT-CHECK CIRCUIT CONNECTIONS” and trigger an
                 alarm.