2. Power On Checks
            a. Turn “MODE” knob (1) to desired setting (A/C, SIMV, CPAP)













                    i. The ventilator will run a SELF-TEST upon set up.  CAUTION: SELF-CHECK must be
                    performed with the disposable ventilator circuit disconnected. Ignoring this requirement
                    could cause the SELF-CHECK process to sense a residual airway pressure leading to a SELF-
                    CHECK failure.

                         1. At this point, CAL is not required. If SELF-TEST results in a Calibration
                         Failure, place (1) to CAL until CAL OK is displayed. If calibration fails,
                         ventilator is deadlined.