P. 206
2. Power On Checks
a. Turn “MODE” knob (1) to desired setting (A/C, SIMV, CPAP)
i. The ventilator will run a SELF-TEST upon set up. CAUTION: SELF-CHECK must be
performed with the disposable ventilator circuit disconnected. Ignoring this requirement
could cause the SELF-CHECK process to sense a residual airway pressure leading to a SELF-
CHECK failure.
1. At this point, CAL is not required. If SELF-TEST results in a Calibration
Failure, place (1) to CAL until CAL OK is displayed. If calibration fails,
ventilator is deadlined.