Weekly Inspections
*ALL CAUTIONS, WARNINGS, AND NOTIFICATIONS THAT CORRESPOND WITH
THE 754 SCREEN WILL BE IN ALL CAPS AND HIGHLIGHTED YELLOW*
1. Complete Duty Inspection
2. Set FiO 2 to 21%
3. Attach vent circuit to vent and field expedient training lung:
a. Slide one large exam glove inside another.
b. Wrap open end of the gloves around the patient end of the circuit tube.
c. Secure with rubber band (DD1380) / Cut-off wrist bead of another glove / Penrose drain / Tape
4. Connect high pressure oxygen source and turn on O 2
5. Turn vent on and allow respirations to begin (listen for compressor)
6. Set FiO 2 to 100%
a. Internal compressor will stop (audible)
7. Turn off oxygen source
a. Ventilator will alarm and show
b.
“O 2 LOW/FAIL-CHECK OXYGEN
SOURCE/CONNECTIONS”
and
“FiO 2 -GAS MIX ERROR. CHECK
SOURCE/SETTINGS/CONNECTIONS”
c. Compressor will turn on (audible)
208