Weekly Inspections
                  *ALL CAUTIONS, WARNINGS, AND NOTIFICATIONS THAT CORRESPOND WITH
                    THE 754 SCREEN WILL BE IN ALL CAPS AND HIGHLIGHTED YELLOW*
            1. Complete Duty Inspection
            2. Set FiO 2  to 21%
            3. Attach vent circuit to vent and field expedient training lung:
                 a. Slide one large exam glove inside another.
                 b. Wrap open end of the gloves around the patient end of the circuit tube.
                 c. Secure with rubber band (DD1380) / Cut-off wrist bead of another glove / Penrose drain / Tape















            4. Connect high pressure oxygen source and turn on O 2
            5. Turn vent on and allow respirations to begin (listen for compressor)
            6. Set FiO 2  to 100%
                 a. Internal compressor will stop (audible)
            7. Turn off oxygen source
                 a. Ventilator will alarm and show
                 b.
                                “O 2  LOW/FAIL-CHECK OXYGEN
                                  SOURCE/CONNECTIONS”
                                          and
                                   “FiO 2 -GAS MIX ERROR. CHECK
                                  SOURCE/SETTINGS/CONNECTIONS”
                 c.  Compressor will turn on (audible)