8. Set FiO 2  to 21%.
          9. Set HIGH pressure alarm to 5cmH 2 O below PEAK.
               a. Ventilator will “stutter” and show “HIGH PRESSURE-PEAK INSPIRATORY PRESSURE TOO
                 HIGH” warning and signal an alarm.
















          10. Turn PLATEAU pressure ON

               a. Ventilator will display “HIGH PRESSURE-PEAK INSPIRATORY PRESSURE TOO HIGH” and
                 “PLATEAU VOLUME-DELIVERED VOLUME LESS THAN SET VOLUME” and trigger an
                 alarm.