8. Set FiO 2 to 21%.
9. Set HIGH pressure alarm to 5cmH 2 O below PEAK.
a. Ventilator will “stutter” and show “HIGH PRESSURE-PEAK INSPIRATORY PRESSURE TOO
HIGH” warning and signal an alarm.
10. Turn PLATEAU pressure ON
a. Ventilator will display “HIGH PRESSURE-PEAK INSPIRATORY PRESSURE TOO HIGH” and
“PLATEAU VOLUME-DELIVERED VOLUME LESS THAN SET VOLUME” and trigger an
alarm.