Page 214 - 2023 SMOG Digital
P. 214

3. Turn FiO 2  knob (#5) to 100%
            4. Cycle back on and to desired settings leaving FiO 2  at 100%
                  a. PEEP will have to be reset when vent is cycled on.

                Note: This technique will transition the ventilator to using oxygen pressure instead of the
                        compressor to drive ventilation and may hasten oxygen usage
                                    Battery Failure
            1. Place patient on BVM with supplemental O 2
            2. Turn ventilator OFF
            3. Replace ventilator battery with battery from 326M suction apparatus. (Per the
               manufacturer, they are the exact same!)
                 a. 326M battery is in the same location as the 754.
            4. Resume normal operations
                 a. PEEP will have to be reset when vent is cycled on.





























                                        Fail Code 6
                   Suctioning to long with inline suction and/or patient inspiratory effort is significant
                   enough to trigger.



        214
   209   210   211   212   213   214   215   216   217   218   219