Page 15 - JSOM Spring 2024
P. 15

Pressures were recorded every second and marked at audible   artery or posterior tibial artery for thighs) was monitored with
              pulse gone and return.                             audible  Doppler  (Ultrasonic  Doppler  Flow  Detector  Model
                                                                 811 with 9.5MHz adult flat probe; Parks Medical Electronics,
              Tourniquet Applications                            www.parksmed.com). Pulse gone was defined as loss of the au-
              Each TPT2 application was directly on skin. In each appli-  dible distal Doppler pulse. Pulse return was defined as return
              cation, careful wrapping was done to achieve 100% side-  of the audible distal Doppler pulse.
              to-side overlap of the inner and outer straps. There was no
              calibrated method of outer strap slack removal when pulling   Once the TPT2 was in place, the stopcocks were used to allow
              strap through the clipped triglide, but outer strap pull-related   the following sequence: 10 seconds with the system at atmo-
              folding of the inner strap at the attachment of the clip was   spheric pressure, TPT2 inflation to just past pulse gone for five
              avoided.                                           pulse gone/return cycles, opening the system to atmospheric
                                                                 pressure for TPT2 deflation, a second TPT2 inflation with a
              Unless there was a reason not to use right limbs, applications   second set of five pulse gone/returns, and opening the system
              were to right distal forearms and right mid-thighs. Subjects   to atmospheric pressure for TPT2 deflation before TPT2 re-
              were seated. Forearm applications had the tourniqueted fore-  moval. A 20mL syringe was used for inflation and pressure
              arm resting on a table. Thigh applications had the knees flexed   cycling of forearm applications, and a 60mL syringe was used
              approximately 90 degrees with both feet resting on the floor.   for thigh applications. Example pressure traces are shown in
              Distal arterial flow (radial artery for forearms, dorsal pedal   Figure 2.
              FIGURE 2  Example forearm and thigh pressure traces.

              (A)                                                (B)
                  200                                                550
                  180                                                500
                 (mmHg)  160                                        (mmHg)  450

                                                                     400
                  140
                 Pressure  120                                      Pressure  350
                                                                     300
                  100
                 TPT2  80                                           TPT2  250
                                                                     200
                 Forearm  60                                        Thigh  150
                   40
                                                                     100
                   20                                                 50

                    0                                                  0
                     0      50    100    150    200    250              0      50    100    150    200    250
                                  Time (seconds)                                     Time (seconds)



              (C)
                  1300
                                 air leak past syringe plunger
                  1200
                  1100
                 (mmHg)  1000
                   900
                 Pressure  800                                   Red circles indicate pulse gones. Blue circles indicate pulse returns.
                                                                 (A) Example forearm TPT2 pressure trace from subject with forearm
                   700
                                                                 circumference  of  14.8cm  and systolic  blood  pressure  of  112mmHg
                                                                 (same subject as panel B). (B) Example occlusive thigh TPT2 pressure
                   600
                 TPT2  500                                       trace from subject with thigh circumference of 53.0cm and systolic
                                                                 blood pressure of 112mmHg (same subject as panel A). (C) Example
                                                                 not occlusive thigh TPT2 pressure trace from subject with thigh cir-
                   400
                 Thigh  300                                      cumference of 61.6cm and systolic blood pressure of 126mmHg.
                   200
                   100
                     0
                      0          50         100        150
                                   Time (seconds)


                                                                                                  TPT2 Pressures  |  13
   10   11   12   13   14   15   16   17   18   19   20