Page 29 - JSOM Summer 2023
P. 29
TABLE S6 Operational Check and Visual Examination Results for the High Temperature Test
Peak Average Peak Average
Inspiration Inspiration Expiration Expiration
Device (L/min) (L/min) (L/min) (L/min) Damage Description Score
Solus A 110 60.4 –67 –51 — 10
Solus B 115 67.3 –79 –58 — 10
Solus C 118 71.7 –82 –59 — 10
i–gel A 84.8 48.3 –63 –46 — 10
i–gel B 73.8 46.5 –49 –37 Small aesthetic damage 9
i–gel C 73.9 50.3 –56 –44 Small aesthetic damage 9
AuraGain A 84.2 52.4 –71 –49 — 10
AuraGain B 82.7 65 –78 –54 — 10
AuraGain C 93.9 51.3 –64 –49 — 10
King A 67.8 44.1 –54 –40 Rip in cuff 6
King B 101 58.3 –73 –54 — 10
King C 89.1 55.3 –74 –53 — 10
Straight A 108 62.1 –73 –56 More flexible, difficult to insert into the mannequin 7
Straight B 91.9 54.8 –74 –53 Significantly more flexible, difficult to insert into the 6
mannequin
Straight C 110 62.3 –81 –59 — 10
air–Q A 81.4 63 –60 –44 More flexible, difficult to insert into the mannequin 7
air–Q B 85.7 53.7 –64 –47 More flexible, difficult to insert into the mannequin 7
air–Q C 86.2 52.7 –58 –45 More flexible, difficult to insert into the mannequin 7
Classic A 113 71.9 –85 –61 — 10
Classic B 123 68.3 –88 –62 — 10
Baska A 107 60.4 –53 –38 Difficult to provide sufficient airflow 5
Baska B 109 55.1 –44 –32 Difficult to provide sufficient airflow 5
Baska C 80.2 49.1 –30 Difficult to provide sufficient airflow 5
Supreme A 86.2 51.6 –52 –39 Small resistance from Ambu bag while providing 8
ventilation
Supreme B 72.9 35.4 –27 –23 Small resistance from Ambu bag while providing 8
ventilation
Supreme C 86.9 51.8 –68 –51 — 10
Fastrach A 110 65.1 –79 –57 — 10
Fastrach B 109 66.1 –80 –58 — 10
Fastrach C 100 62.5 –80 –58 — 10
SuperNova A 119 54 –75 –55 — 10
SuperNova B 83.5 45.9 –59 –46 Silicone plug became loose 7
SuperNova C 79.2 53 –69 –51 — 10
TABLE S7 Operational Check and Visual Examination Results for the Low Temperature Test
Peak Average Peak Average
Inspiration Inspiration Expiration Expiration
Device (L/min) (L/min) (L/min) (L/min) Damage Description Score
Solus A 55.2 35.8 –34 –28 Sufficient airflow cannot be provided 3
Solus B 110 64.5 –73 –52 — 10
Solus C 104 68.9 –81 –58 — 10
i–gel A 99.4 58.8 –66 –51 — 10
i–gel B 98.7 59.8 –67 –52 — 10
i–gel C 94.8 62.3 –66 –48 — 10
AuraGain A 103 71.8 –86 –60 — 10
AuraGain B 111 78.1 –87 –64 — 10
AuraGain C 91.8 49.4 –63 –48 — 10
King A 62.2 35.7 –36 –29 — 10
King B 97.3 68.7 –85 –0.59 — 10
King C 121 73.1 –90 –62 — 10
Straight A 122 68.3 –83 –61 — 10
Straight B 113 69.2 –85 –60 — 10
Straight C 110 67.9 –86 –58 — 10
(continues)
Testing SGA Devices in Military Combat | 27