Page 21 - JSOM Spring 2021
P. 21

units of PRBCs infused simultaneously at the emergent flow   The measurements of interest for this study were mean tem-
              instead of 1L. These tests were done at ground level and at   perature, time to reach mean temperature, change in tempera-
              8,000- and 16,000-ft simulated altitude in an altitude cham-  ture from inlet to outlet, proportion of time the temperature
              ber. The ambient temperature inside the altitude chamber was   was ≥32°C and ≥35°C, and battery life.
              maintained at 24°C to approximate room temperature in our
              laboratory (23.9 ± 0.4°C) in which ground level testing was   Statistical Analysis
              completed. Two of each device were used in the study and two
              tests with each device were completed at each condition. All   Temperature differences at specified conditions, NS and
              devices were operated on battery power, and the Buddy Liter     PRBCs, emergent flow and nonemergent flow rate, and alti-
              and Buddy Lite AC were also operated on alternating current   tude differences were compared for each device. Mean and SD
              (AC) power. The Thermal Angel and M Warmer do not offer   were used to summarize data. Comparisons were made using
              the option of operating from AC power.             the general linear model univariate analysis to create contrasts
                                                                 that tested specified custom hypotheses. This method was
              Battery life was measured under two conditions with each de-  preferred to avoid testing all pairwise comparisons, as com-
              vice: nonemergent flow of 125mL/h using room temperature   parisons were set a priori. Post hoc analysis was completed
              NS as described above and using a pressure bag inflated to   by comparing the change in temperature prewarmer to post-
              300mmHg using iced NS. Each device was operated until the   warmer, and time to reach mean temperature was created with
              low battery indicator was activated and measured temperature   each device at all conditions using the Student t-test. Devices
              decreased by >1°C.                                 were compared to each other at all conditions and compared
                                                                 to themselves using altitude as the independent variable, to
              Devices were set up per manufacturer’s instructions. The in-  determine if altitude had an effect on device performance. Sta-
              fusion pump had preventive maintenance and calibration per-  tistical significance was determined at α = .05, two-tailed, and
              formed before the study began. Device batteries were charged   SPSS  Statistics 25 was used for data analysis.
                                                                    ®
              for a minimum of 24 hours before use. Standard IV tubing was
              used for all nonemergent flows with the infusion pump and   Results
              pressure bag, and standard blood tubing was used for testing
              with PRBCs. A three-way stopcock was placed at the entrance   Table 1 shows the physical and operational characteristics of
              to and directly after the heater unit of the device being tested,   the devices. There was no distinct size or weight advantage of
              and a J-type thermocouple (Omega Engineering, https://www   one device over the others. The stated maximum flows were
              .omega.com/en-us/thermocouple-types) was placed in each of   much higher for the Thermal Angel and M Warmer. None of
              the stopcocks’ open port and sealed with silicone. The thermo-  the devices were able to heat NS or cold PRBCs to a mean
              couples were attached to a data acquisition system (National   temperature equal to normal human body temperature (37°C)
              Instruments,  https://www.ni.com/en-us.html), and  tempera-  at either flow tested. Analysis of the differences in overall tem-
              ture data were continuously recorded at 1-second intervals.   perature profiles produced by the devices at all conditions was
              The Buddy Liter and Buddy Lite use disposable cartridges in-  statistically significant (p < .01) except for the Buddy Liter
              side the reusable heater unit, and the Thermal Angel and M   versus the Buddy Lite at ground level on AC power, at the
              Warmer use disposable heater units. All the cartridges/heater   nonemergent flow using room temperature NS (p = .42), and
              units except the Thermal Angel had tubing before and after   the same devices at 16,000-ft altitude on AC power at the
              the heater units. These are the points at which the preheater   nonemergent flow, using cold PRBCs (p = .24). Figure 2 shows
              and postheater temperatures were measured. A 9-inch IV ex-  the mean (SD) temperatures for all devices on battery power.
              tension tubing supplied with the Thermal Angel was placed   The M Warmer produced the highest mean temperatures at
              on the output side of the heater so that measurements could   the emergent flow using NS and cold PRBCs. Differences in
              be taken in the same location with all devices and would sim-  mean temperature between devices at the nonemergent flow
              ulate the temperature at which the warmed fluid would enter   using both NS and PRBCs were within 3°C at each of the
              a patient’s circulation. After priming, each warmer was turned   altitude conditions, although the temperatures were some-
              on, and fluid flow and temperature measurements were started   what  lower at  altitude  compared  to  ground level.  Figure  3
              simultaneously.                                    shows the mean (SD) temperatures for the Buddy Liter and

              TABLE 1 Physical and Operational Characteristics for Each Fluid Warming Device
                                               Buddy Liter     Buddy Lite AC ™  Thermal Angel      M Warmer
              Dimensions (in) (L ´ W ´ H)
              Battery housing                4.92 ´ 3.33 ´ 1.36  7.26 ´ 3.33 ´ 1.36  6.4 ´ 3.2 ´ 1.7  7.09 ´ 3.54 ´ 1.38
              Heater unit                     5.2 ´ 1.5 ´ 0.87  15.2x 1.5 ´ 0.87  9.0 ´ 2.9 ´ 0.95  3.94 ´ 1.97 ´ 0.79
              Weight (lb)
              Battery and heater unit             1.09             1.46             1.83              1.68
              AC power supply                     2.64             2.64             N/A               0.55
              Power requirement                AC, battery       AC, battery       Battery           Battery
              Temperature set point (°C)         38 ± 2           38 ± 2           38 ± 3            39 ± 3
              High temperature alarm              Yes               Yes           LED only          LED only
              Low temperature/no heat alarm       Yes               Yes           LED only          LED only
                                               30 @ 20°C         80 @ 20°C
              Maximum flow rate (mL/min)                                         150 @ 20°C      150 @ 4°C – 37°C
                                               20 @ 10°C         50 @ 10°C
              L ´ W ´ H = length ´ width ´ height, LED = light-emitting diode.

                                                                                Fluid Warming Technology Performance  |  19
   16   17   18   19   20   21   22   23   24   25   26