Page 51 - Journal of Special Operations Medicine - Summer 2016
P. 51

Disinfectant versus antiseptic: A disinfectant is an agent   Steam Sterilization
              intended to be used on surfaces and equipment, but not   Steam sterilization, known as autoclaving, is the recom-
              skin, as they can cause damage to living tissue. Antisep-  mended method for sterilization of medical equipment
              tics are agents that are intended to be applied to skin   that is both heat and moisture tolerant.  This form of
                                                                                                   1,2
              and tissue but are inappropriate for cleaning equipment   sterilization uses moist heat in the form of steam under
              and surfaces.                                      pressure. It differs from a simple open container boil-
                                                                 ing technique since the addition of pressure allows the
                                                                 autoclave to achieve temperatures beyond the boiling
              Requirements
                                                                 point of water. These high temperatures allow the auto-
              Not all medical equipment needs to be sterilized, but it is   clave to kill microorganisms in a relatively short period
              important to identify whether cleaning, disinfection, or   of time. It is a dependable form of field sterilization;
              sterilization of the item is required based on that item’s   however, its effectiveness is limited by steam, pressure,
              intended  use.   In  ideal  situations,  items  that  come  in   temperature, and time. 8
                         8
              contact with unbroken skin (stethoscopes, blood pres-
              sure cuffs) should be cleaned with low-level disinfection.   An autoclave must be able to achieve and maintain a
              Items that touch but do not penetrate mucus membranes   temperature of 121°C (250°F) or 132°C (270°F) to have
              (endotracheal tubes, laryngoscopes) should be sterilized   affective antimicrobial activity.
              but can be cleaned with high-level disinfection if steril-
              ization is not feasible. Medical instruments that pierce   There is disparity over the proper exposure time re-
              human tissue (scalpels, forceps,  cricothyroid hooks)   quired to affectively sterilize equipment. The CDC
              must be sterilized between each patient contact. 1,7,8  Bear   and the Special Operations Forces Medical Handbook
              in mind that this is the recommendation for fixed hos-  recommend 30 minutes at 121°C in a gravity displace-
              pitals in a garrison setting. The operational setting is   ment autoclave or 4 minutes at 132°C in a prevacuum
              dictated by the tactical and logistical limitations, while   sterilizer. 8,13  However, according to the WHO, a mini-
              simultaneously conforming to some semblance of a pol-  mum duration of 15 minutes at a temperature of 121°C
              icy for safe instrument reuse.                     (250°F) seems to be an appropriate cleaning time. If a
                                                                 formal autoclave is not available, any commercial pres-
              Simply put, cleaning and sterilization/disinfection of in-  sure cooker can suffice. 8
              struments represent a process that makes them safe for
              medical personnel to handle and safe to reuse on a pa-  The autoclave can potentiate logistical constraints since
              tient.  Figure 1 shows a suggested instrument processing   it requires a large amount of water to operate. Any
                  6
              cycle.                                             source of potable water can be used, but nondistilled
                                                                 water can cause ionized salts like calcium, magnesium,
                                                                 and iron to precipitate and deposit on lower points of the
              Sterilizers                                        autoclave. For this reason, most autoclave manufactur-
                                                                 ers recommend the use of distilled water for operation.
              Sterilization                                      In the garrison hospital or clinic setting, this is usually
              Several methods are available to achieve instrument   not a problem. In the field environment, however, large
              sterilization; however, most are not accessible in the op-  quantities of distilled water are not available. For this
              erational environment. Therefore, we will focus on the   reason, the US Army Medical Materiel Agency National
              two forms that are likely to be available in the deployed   Maintenance Program recommends that local potable
              setting: steam sterilization and dry heat sterilization.  water may be used for daily operations; however, the
                                                                 unit should be cleaned periodically with a diluted vin-
              Figure 1  Sample decontamination process.          egar and water solution (1 quart vinegar:4 gallons of
                                                                 water) to remove any ion salt buildup.  If your kit in-
                                                                                                   16
                                                                 cludes an electrical autoclave, then the electrical heating
                                                                 element should be cleaned in a similar fashion, as ion
                                                                 salts can deposit there as well.

                                                                 Another logistical  constraint of the sterilizer  is its
                                                                 weight. Depending on its size, a standard unit can weigh
                                                                 between 11 and 43lb. Obviously, smaller units weight
                                                                 less, but this comes as a tradeoff for its reduced instru-
                                                                 ment capacity. Nevertheless, the steam sterilizer’s size
                                                                 and weight make it ill-suited for a portable devise and is
                                                                 best used at the Role 1 facility or higher.



              Field Sterilization in the Austere Environment                                                  37
   46   47   48   49   50   51   52   53   54   55   56