Page 29 - JSOM Fall 2025
P. 29

Two factors that affect the time it takes for fluid to flow   temperature, one measures the IV tubing inlet temperature, two
              through the IV tubing are flow rate and tubing length. Bisson-  measure the outlet temperature, and two gauge the chamber’s
              nette and Paut found that low fluid flow rates caused by low   environmental temperature.  The thermocouples have an SD
              environmental temperature  can aggravate heat loss through   of 2.2°C or 2% accuracy between –50 and 99°C. A Transonic
              IV tubing.  Tubing length has effects on the time required for   ME-5PXL clamp-on flowmeter (Transonic, Ithaca, NY) near
                      10
              fluid to flow through the IV tubing. Holt identifies IV tubing   the IV tubing outlet measures the volumetric flow rate with SD
              as a potential source of heat loss in his study which presents   8.0mL/min accuracy.  To quantify the uncertainty associated
                                                                                 13
              evidence that indicates that shorter IV tubing at the output of   with heat loss, the Kline-McClintock method is employed.
              inline warming devices produces less heat loss.  Greater heat
                                                  11
              loss could cause the fluid temperature to decrease from healthy   Experimental measurements were conducted by systematically
              temperatures to unhealthy.                         reducing the environmental chamber temperature in 3°C incre-
                                                                 ments from 20°C to –39°C, while maintaining the inlet fluid
              This research tests the dual hypothesis that (1) heat loss   temperature at 38°C (SD 3.3°C). At each temperature setting,
              through IV tubing in field blood transfusion kits increases sig-  data were collected from all thermocouples and the flow meter.
              nificantly with decreasing environmental temperatures, and   The collected measurements included fluid temperatures at
              (2) modifications to tubing insulation and length can effec-  both the inlet and outlet of the IV tubing and the volumetric
              tively reduce this heat loss.                      flow rate of the glycerol solution. Heat loss through the IV
                                                                 tubing was then calculated by using the measured data.
              Methods
                                                                 According to Frinak et al.,  the use of 22% glycerol by weight
                                                                                     14
              A portable in vitro testing apparatus was developed to test   in water with a specific gravity of 1.055 at 20°C is confirmed
              the FWB transfusion kit in an environmental chamber with   as an appropriate blood simulant that provides viscosity char-
              an interior volume of 1.81m³. The chamber operates within   acteristics similar to whole blood. Hence, a 22% by mass
              a temperature range of –80 to 200°C and an altitude range   aqueous glycerol solution is used in experiments to mimic the
              of 244 to 30,480m. The testing apparatus was built using an   fluid properties of FWB. This solution has a density of 1016–
              aluminum framing rail and includes two reservoirs: one on   1020kg/m³ and a specific heat capacity of 3913–3914J/(kg∙K),
              the ground and one at a desired height. The upper reservoir   which means its density is 0.88%–3.78% lower and its spe-
              maintains a constant pressure head for fluid flow through IV   cific heat capacity is 0.67%–8.39% higher than FWB. Since
              tubing using an overflow line that redirects excess fluid back   glycerol has a higher density and lower specific heat capacity
              to the lower reservoir. To keep the fluid level steady in the up-  than water, increasing the glycerol percentage would further
              per reservoir, a pump transfers excess fluid from the lower res-  align the solution’s properties with those of FWB.
              ervoir back to the upper one. An acrylic platform supports the
              insulated upper reservoir, which contains an immersion heater   Results
              to control the fluid temperature at the IV tubing’s inlet to 38°C
              (SD 3.3°C), simulating healthy FWB temperature for trauma   Heat Loss Analysis
              patients. The IV tubing runs from the bottom of the insulated   Inlet and outlet fluid temperatures are measured using the
              reservoir to a collection reservoir at the bottom of the frame.   thermocouples located at the inlet and outlet of the IV tubing
              The height differential between the outlet of the IV tubing and   component of the FWB transfusion kit. During experiments,
              the overflow is 440mm.                             the temperature within the environmental chamber is system-
                                                                 atically reduced by 3°C from the first measurement starting at
              The field FWB transfusion kits have several components in-  20°C to a final one at –39°C. Based on the cold weather test-
              cluding a 450mL blood collection bag, IV tubing, anticoagu-  ing standards outlined in the U.S. Army Test and Evaluation
              lant citrate phosphate dextrose solution, safety IV catheters,   Command (ATEC), MIL-STD-810H, which are used to qual-
              needle port, etc. The Fresh Whole Blood Recipient Set (Safe-  ify medical equipment for arctic deployment, the lower tem-
              guard Medical, Huntersville, NC), purchased from the man-  perature limit was chosen as –39°C.  The upper limit of 20°C
                                                                                            15
              ufacturer) was used in our testing protocol. The IV tubing set   serves as a control. The temperature of the fluid at the inlet of
              “Y-type” with a filter, shown in the middle of Figure 1,  is the   the IV tubing is 38°C (SD 3.3°C). The measurements from the
                                                        12
              focus of this study and the length of the tubing is 203cm.   eight thermocouples and ultrasonic flowmeter are recorded at
                                                                 each setting of the environmental temperature. Figure 2 pres-
                                                                 ents the thermocouple measurements of fluid temperatures at
              FIGURE 1  Testing apparatus.
                                                                 the IV tubing inlet and outlet across the environmental tem-
                                                                 perature range of –39°C to 20°C.
                                                                 Figure 2 illustrates a significant temperature drop in the glyc-
                                                                 erol solution from 39.2 to 34.1°C, even at an environmental
                                                                 temperature of 20°C. The temperature differential between the
                                                                 inlet and outlet increases linearly as the environmental tem-
                                                                 perature decreases. At 0°C, a 30% temperature reduction is
                                                                 observed. When the environmental temperature decreases to
                                                                 –39°C, the outlet temperature of the glycerol solution drops
                                                                 to 9.1°C (SD 0.35°C), representing a 75.7% decrease. While
              The testing apparatus uses eight K-type thermocouples and an   glycerol solution can still be delivered through the IV tubing at
              ultrasonic flowmeter to measure temperatures and flow rates.   –39°C, this substantial temperature reduction may pose risks
              Three thermocouples monitor the insulated reservoir’s fluid   to the patient’s core body temperature.

                                                                      Mitigating FWB Heat Loss During Austere Transfusions  |  27
   24   25   26   27   28   29   30   31   32   33   34