Page 38 - JSOM Summer 2024
P. 38

Results
          FiO 2
          FiO  during use of the 731 was not impacted by ventilator   FIGURE 6
                                                             Effect of
             2
          settings (p>.99). Mean FiO was 0.96 (SD 0.25) for all set-  ventilator settings
                                2
          tings. FiO was significantly (p=.04) associated with altitude   on humidity
                  2
          and  tended  to be  lower  at higher  altitudes  although  differ-  (mean and range).
          ences were not clinically important. Mean FiO was 0.97 (SD

                                              2
          0.02) at ground level, 0.96 (SD 0.02) at 8,000 feet, and 0.95
          (SD 0.03) at 16,000 feet (Figure 3). FiO was significantly
                                            2
          (p<.001) higher with 3L/min oxygen bleed-in, versus 1L/min.   Temperature and Humidity
          Mean FiO was 0.94 (SD 0.02) with 1L/min oxygen bleed-in
                  2
          and 0.98 (SD 0.01) with 3L/min oxygen bleed-in (Figure 4).   Mean  humidity  was  89%  (range  87%–90%)  for  RR30/V
                                                                                                            T
          Minimum FiO in each group was 0.92 and 0.97, respectively.  250 settings, 92% (91%–93%) for RR20/V  450, and 96%
                     2                                                                          T
                                                             (95%–97%) for RR20/V  700 (Figure 6). Differences were
                                                                                 T
                                                             statistically significant overall (p<.0001) as were all pairwise
                                                             comparisons. Peak temperature did not differ significantly
                                                             between  the  three  RR/V   ventilator  settings  (p=.47).  Group
                                                                                 T
                                 FIGURE 3  Effect of Altitude on   means were 25.5°C, 25.2°C, and 25.6°C, respectively.
                                 Fio2 (mean and range) showing
                                 combined 1 and 3L/min oxygen
                                 flows at each altitude with the    CO  Absorbent Life
                                                                2
                                 731 ventilator.             Mean time to CO  elevation of >1% was 456 (SD 56) mins
                                                                           2
                                                             for RR30/V  250, 460 (SD 44) mins for RR20/V  450, and
                                                                      T
                                                                                                     T
                                                             557 (SD 27) mins for RR20/V  700 ventilator settings (Figure
                                                                                    T
                                                             7). This difference was statistically significant overall (p=.005)
                                                             but not all pairs were significantly different: RR20/V  700 was
                                                                                                     T
                                                             significantly different from the other ventilator settings (both
                                                             p<.01), but the RR30/V  250 and RR20/V  450 settings were
                                                                                              T
                                                                                T
                                                             not significantly different from each other (p=.89.)
          FIGURE 4  Effect of
          oxygen flow on Fio
                        2
          (mean and range) with
          the 731 ventilator.                                                                     FIGURE 7
                                                                                                  Effect of
                                                                                                  ventilator
                                                                                                  settings on
                                                                                                  time to CO
                                                                                                          2
                                                                                                  >1% (mean
          FiO  during use of the SAVe II was not associated with ventila-                         and range).
             2
          tor settings (p=.73) or altitude (p=.97). It should be noted that
          the maximum PEEP setting with the SAVe II was 10cm H O,
                                                         11
                                                      2
          therefore the ventilator settings requiring 20cm H O PEEP   Condensate from the ventilator circuit was drained and mea-
                                                   2
          were not possible. There was a minimal decrease in mean FiO    sured after  each test (data  not shown). The longest the  run
                                                         2
          with increasing altitude: 0.95 (SD 0.03) at ground level, 0.94   times produced the largest condensate volume. The conden-
          (SD 0.26) at 8,000 feet, and 0.94 (SD 0.27) at 16,000 feet. FiO 2   sate volume produced by the RR30/V  250 and RR20/V  450
          was significantly (p<.0001) higher with 3L/min oxygen, versus   ventilator  settings  run times  were  not significantly  different
                                                                                          T
                                                                                                         T
          1L/min and was less variable (Figure 5). Mean FiO was 0.92     (p=.55, range 25–35mL) whereas these settings produced con-
                                                  2
          (SD 0.01) at 1L/min oxygen, and 0.97 (SD 0.01) at 3L/min   densate volumes that were significantly different from volumes
          (mean difference 0.05). Minimum FiO in each group was   produced using the RR20/V  700 ventilator settings (range
                                         2
          0.9 and 0.96, respectively. Due to the limited flow limita-  45–63mL, p=.005 and p=.0003, respectively).
                                                                                    T
          tions (24L/min)  and PEEP limitations of the SAVe II, direct
                      11
          FiO  comparisons to the 731 could only be made using the   Suctioning and Disconnect
             2
          RR30/V  250/PEEP 5 settings. Differences in FiO  between the   FiO  at baseline, after 5 and 10 seconds of airway suctioning,
                                                2
                T
          two devices at these settings were not significantly different     and after 15 and 30 seconds of a ventilator disconnect, and re-
                                                                2
          (p=.97).
                                                             covery times back to baseline FiO  are shown in Table 3. With
                                                                                       2
                                                             the exception of the 15-second ventilator circuit disconnect
                                                             and the 5-second suction recovery time, the 1L/min O  bleed-in
                                                                                                      2
                                                             tests resulted in significantly greater decreases in FiO  and sig-
                                                                                                      2
                                                             nificantly longer recovery time than the 3L/min tests and at
                                       FIGURE 5  Effect of
                                       oxygen flow on Fio2 (mean   baseline (p<.05). The 15-second ventilator circuit disconnect
                                       and range) with the SAVe II   did not result in significantly different FiO  with either the 1 or
                                                                                             2
                                       ventilator.           3L/min O  tests (p>.1).  Within-group  (1 and 3L/min)  FiO
                                                                     2
                                                                                                            2
                                                             and recovery  time was  significantly  different  (p<.05)  with
                                                             the exception of the 15-second ventilator circuit disconnect
                                                             (p>.1).
          36  |  JSOM   Volume 24, Edition 2 / Summer 2024
   33   34   35   36   37   38   39   40   41   42   43