Page 38 - JSOM Spring 2026
P. 38

FIGURE 2  Example pressure traces.                 analyses were performed with GraphPad Prism, version 7.04
                                                             for  Windows (GraphPad Software Inc., Boston, MA, USA;
                                                             www.graphpad.com). Unpaired t tests were used for pressure,
                                                             tightening-system use, and pressure/click comparisons. Paired
                                                             t tests were used for per-click pressure comparisons for first-
                                                             completion versus after-first-completion. Pearson’s correlation
                                                             was used for circumference versus occlusion pressure and cir-
                                                             cumference and pre-tightening-system use versus tightening-
                                                             system use. Fisher’s exact test was used for contingency tables.


                                                             Results
                                                             Appliers
                                                             There were 7 appliers (1 male, 6 females, 6 undergraduates,
                                                             1 team leader). All of the undergraduate appliers were 19 to
                                                             22 years old. Three of the 6 undergraduate approved appliers
                                                             were  in their first year  with the research  team  (appliers on
                                                             14, 9, and 6 recipients). Two were in their second year with
                                                             the team (appliers on 4 and 2 recipients); the tourniquet with
                                                             which they gained experience in the prior year was the Tac-
          Example traces of tourniquet pressures from application through re-  tical Pneumatic Tourniquet 2inch (Alphapointe  Tactical  &
                                                                                                   ™
          lease and removal. Pressures were collected every second and shown
          as small open circles with connecting lines. Named pressure and time   Medical, Kansas City, MO, USA; Alphapointe.com). One stu-
          events are each marked by an × pointed at by an arrow from a text la-  dent was in her third year (applier on 4 recipients); so she had
          bel indicating the name of the event. The arm application trace is green.   experience with the Tactical Pneumatic Tourniquet 2inch and
          The thigh application trace is blue. The limb circumferences are shown   the OMNA Marine tourniquet (OMNA Inc. St. Petersburg,
          with each trace. The arm application did not have a return of arterial
          flow before release and therefore also did not have a 2nd completion.  FL, USA; www.marinetourniquets.com). The team leader who
                                                             applied on 1 recipient was 59 years old.
          X8T happened after each advancement click. First completion
          was one click past first arterial occlusion. Returns of arterial   Recipients
          flow before 100 seconds from first completion resulted in an   Table 1 has recipient information.
          additional tightening click. Pre-release pressure was taken 100
          seconds from first completion.                     Pressures and Tightening-System Use
                                                             All 80 X8T applications reached completion. Five arm and 28
          PW supervised applications and called Doppler signal loss and   thigh applications had pulse return before 100 seconds requir-
          return. A separate person ran the pressure collection software   ing additional tightening (Figure 3).
          and also listened for Doppler signal loss and return. Recipients
          had the option to stop at any time.                Six arm applications occluded without tightening-system use.
                                                             Four of those did not require tightening-system use beyond the
          Resetting and Cleaning                             click to completion. After first completion, three arm applica-
          The dial and strap/redirect systems were reset after each use.   tions required an additional click once and two required an
          Each X8T was sprayed with 70% denatured ethyl alcohol and   additional click twice (Figure 3). X8T arm application pres-
          allowed to dry between recipients.                 sures are only for arms >31.2cm circumference, but clicks were
                                                             counted for every application.
          Statistical Analysis
          Data were organized in Excel  for Microsoft 365 MSO ( Version   No thigh  applications occluded  without tightening-system
                                ®
          2508 Build 16.0.19127.20082 64-bit; Microsoft Corp., Red-  use (P=.026 versus arm). After first completion, 25 thigh ap-
          mond, WA, USA; www.microsoft.com). Graphing and statistical   plications required an additional click once, two required an


          TABLE 1  Tourniquet Recipient Information
                                               Group; median (minimum, 25th percentile, 75th percentile, maximum)
                                              All             Arm circumference <31.3cm  Arm circumference ≥31.3cm
           Number of recipients               40                       21                       19
           Sex, F; M                        27 F, 13 M               19 F, 2 M               8 F, 11 M
           Age, y                        20 (19, 20, 22, 68)     20 (19, 20, 20.5, 64)    21 (19, 20, 58, 68)
           Height, cm                  170 (152, 163, 180, 193)  165 (152, 161, 172, 191)  180 (155, 168, 183, 193)
           Weight, kg                71.4 (47.2, 62.8, 95.2, 133.8)  63.5 (47.2, 56.3, 69.2, 81.7)  95.3 (68.0, 77.1, 102.1, 133.8)
           Systolic blood pressure, mmHg  124 (91, 113, 131, 165)  114 (91, 110, 126, 130)  131 (112, 122, 139, 165)
           Arm circumference, cm      31.3 (23.1, 26.7, 35.4, 40.5)  26.8 (23.1, 25.3, 29.6, 31.5)  35.4 (32.0, 32.9, 37.0, 40.5)
                                    [n=19 arm X8T pressures group
                                     35.4 (32.0, 32.9, 37.0, 40.5)]
           Thigh circumference, cm    56.1 (43.7, 52.5, 61.5, 75.5)  52.8 (43.7, 49.8, 55.8, 62.1)  61.2 (54.3, 56.1, 63.8, 75.5)
          X8T = X8T-T2G tourniquet.

          36  |  JSOM   Volume 26, Edition 1 / Spring 2026
   33   34   35   36   37   38   39   40   41   42   43