are uncommon. We demonstrated that an oxygen concentra- Disclosures
tion above 45% was required, which is only found in very Dr. Ho serves as the medical director to Axon Enterprises, Inc.
close proximity to a maximally flowing oxygen source. This and owns shares of stock in the company. Dr. Dawes discloses
supports the concept of rapid oxygen concentration diffusion that he serves as an expert medical consultant to Axon Enter-
during supplemental delivery, even with maximum flow rates. prises, Inc. Dr. Driver has no conflicts to declare.
We acknowledge that CEW use in general is not without some
risk, and we advocate limiting sources of ignition in the patient Funding
care environment for general safety reasons. However, we also This work was supported by Axon Enterprise, Inc. who pro-
believe that prohibiting CEW presence or use because of con- vided lab space, commercial product for study, and engineer-
cerns regarding the use of supplemental oxygen is unfounded. ing assistance during the data gathering phase of this study.
Limitations References
We did not examine every type of TEMS patient care setting 1. Kansagra SM, Rao SR, Sullivan AF, et al. A survey of work-
possible. However, we elected to examine oxygen levels using place violence across 65 U.S. emergency departments. Acad
a space that could be fully enclosed to theoretically limit the Emerg Med. 2008;15(12):1268–1274. doi:10.1111/j.1553-2712.
2008.00282.x
diffusion of the supplemental oxygen. We believe that the data 2. Nachreiner NM, Gerberich SG, Ryan AD, McGovern PM. Minne-
we gathered represents the environment most likely to have sota nurses’ study: perceptions of violence and the work environ-
the highest oxygen concentrations, which would be analogous ment. Ind Health. 2007;45(5):672–678. doi:10.2486/indhealth.
to a worst-case scenario for a CEW arc to occur from a fire- 45.672
risk perspective. 3. Barlow CB, Rizzo AG. Violence against surgical residents. West J
Med. 1997;167(2):74–78.
4. Lavoie FW, Carter GL, Danzl DF, Berg RL. Emergency department
Additionally, we did not test every conceivable combustible violence in United States teaching hospitals. Ann Emerg Med.
material that could be present. We chose materials that would 1988;17(11):1227–1233. doi:10.1016/s0196-0644(88)80076-3
have high likelihood of being in very close proximity to a flow- 5. National Institute for Occupational Safety and Health. Violence
ing oxygen source in a patient care setting. There may be other Occupational Hazards in Hospitals. Centers for Disease Con-
combustible materials that might have lower ignition thresh- trol and Prevention; 2002. Accessed July 6, 2023. https://www.
olds than what we tested. In a TEMS setting, caution needs to cdc.gov/niosh/docs/2002-101/pdfs/2002-101.pdf?id=10.26616/
NIOSHPUB2002101
be used when operating a CEW in an environment of elevated 6. Kowalenko T, Gates D, Gillespie GL, Succop P, Mentzel TK. Pro-
flammability (e.g., in proximity to gasoline vapor). spective study of violence against ED workers. Am J Emerg Med.
2013;31(1):197–205. doi:10.1016/j.ajem.2012.07.010
Finally, we did not test every oxygen concentration level possi- 7. Ho JD, Clinton JE, Lappe MA, Heegaard WG, Williams MF, Miner
ble. We found human hair ignited quite easily at 45% oxygen JR. Introduction of the conducted electrical weapon into a hos-
concentration but did not do so at 35%. We acknowledge that pital setting. J Emerg Med. 2011;41(3):317–323. doi:10.1016/j.
we do not know the exact oxygen concentration between 35% jemermed.2009.09.031
and 45% at which ignition would occur. However, we were 8. Bulman P. Police Use of Force: The Impact of Less-Lethal Weap-
ons and Tactics. National Institute of Justice Journal. March 2,
unable to reproduce conditions in our baseline measurements 2011. Accessed December 12, 2023. https://nij.ojp.gov/topics/
of an enclosed space that could reliably give oxygen concen- articles/police-use-force-impact-less-lethal-weapons-and-tactics
trations above 31.5%, so we believe it to be unlikely that a 9. Kroll MW, Ritter MB, Williams HE. Fatal and non-fatal burn in-
CEW would be used in a field patient care situation where an juries with electrical weapons and explosive fumes. J Forensic Leg
oxygen concentration would be greater than 35%. Med. 2017;50:6–11. doi:10.1016/j.jflm.2017.06.001
10. Kim M-S, Lee J-H, Lee D-H, Lee YU, Jung T-E. Electrocautery-
ignited surgical field fire caused by a high oxygen level during
Conclusion tracheostomy. Korean J Thorac Cardiovasc Surg. 2014;47(5): 491–
493. doi:10.5090/kjtcs.2014.47.5.491
This work represents the first study to evaluate the ignition risk 11. Merriam-Webster Online Dictionary. Merriam-Webster, Inc.;
of CEW use in an environment where supplemental oxygen 2023. “Paschen’s Law.” Accessed July 6, 2023. https://www.
is used. These findings are applicable to patient care settings, merriam-webster.com/dictionary/Paschen%27s%20law#
especially those where TEMS operators work. The risk of fire 12. CE Baukal, Jr. Oxygen-Enhanced Combustion, 2nd ed. CRC
Press; 2013.
ignition from a CEW in this setting is sufficiently remote to 13. Axon Enterprise, Inc. TASER Handheld Energy Weapon Warn-
safely allow use when necessary by a properly trained person. ings. September 20, 2022. Accessed July 5, 2023. https://a.story
Further studies should be undertaken to better understand the blok.com/f/198504/x/1a1dc1606c/law-enforcement-warnings
safety of these devices when used in other environments. -8_5x11.pdf
14. Bigham BL, Jensen JL, Tavares W, et al. Paramedic self-reported ex-
Acknowledgments posure to violence in the emergency medical services (EMS) work-
The authors would like to acknowledge the engineering con- place: a mixed-methods cross-sectional survey. Prehosp Emerg
Care. 2014;18(4):489–494. doi:10.3109/10903127.2014.912703
tributions of Matt Carver and Erinn McIntyre for their help in 15. Grange JT, Corbett SW. Violence against emergency medical ser-
constructing and carrying out the study. vices personnel. Prehosp Emerg Care. 2002;6(2):186–190. doi:
10.1080/10903120290938526
Author Contributions
JH conceived the study concept and obtained funding. JH, DD, PMID: 38865654; DOI: 10.55460/BFZE-IKTR
and BD coordinated and collected the data, and BD analyzed
the data. JH wrote the first draft, and all authors read and
approved the final manuscript.
CEW Fire Risk with Supplemental Oxygen | 43