APPENDIX A.
            APPLICATION-STEP-RELATED TIMES FROM DISPLAYED TIMER



          Times Points                                            with additional attempts to pull the strap tighter mixed
          1.  “Go”                                                in with use of the tightening system, “touch tightening
          2.  Strap/redirect secured                              system” is the first time at which applier touches tight-
          3.  Touch tightening system                             ening system as part of starting to use tightening system.
          4.  First Doppler signal loss                           If the holding location includes part of the tightening
          5.  Last Doppler signal loss before hands off           system, then the time to touch tightening system is to
          6.  Hands off for Done                                  when applier starts to use the tightening system.
                                                               c.  If applier puts the tourniquet on the limb and works on
          Instructions Regarding Time Points                      securing and on tightening but then undoes the tour-
          1.  Start of “Go”                                       niquet and restarts the securing process, then “touch
          2.  Strap/redirect secured                              tightening system” is not  the very first  time at which
            a.  If applier only puts the tourniquet on the limb once and   applier touches tightening system as part of starting to
               goes from placing the tourniquet around the limb to   use tightening system. Instead, “touch tightening sys-
               securing the tourniquet to using the tightening system   tem” is the “touch tightening system” time within the
               with additional attempts to pull the strap tighter mixed   last set of secured to tightening system use. “List other”
               in with use of the tightening system, “strap/redirect se-  in the “Problems with or without failure” receives the
               cured” is first time at which applier stops securing strap   explanation.
               to go to tightening system. If applier continues to pull   4.  First Doppler signal loss
               on the strap while using the tightening system, time of   a.  With ratcheting buckle tourniquets, this is the time with
               “strap/redirect secured” will match time of “touch tight-  no pulse with the buckle in its flat resting position. With
               ening system.”                                     tightening systems that rotate in a plane parallel to the
            b.  If applier never secures strap/redirect system, no other   limb (windlass and X8T), this is the time of the missing
               times recorded.                                    beat that is the first of the missing beats of occlusion. If
            c.  If applier puts the tourniquet on the limb and works on   the applier fails to achieve a Doppler signal loss, there
               securing or even on tightening but then undoes the tour-  will not be a time.
               niquet and restarts the securing process, then “strap/  5.  Last Doppler signal loss before hands off
               redirect secured” is not the very first time at which ap-  a.  For applications with only one Doppler signal loss, this
               plier stops securing the strap. Instead, “strap/redirect   is the same time as first Doppler signal loss.
               secured” is the “strap/redirect secured” time within the   b.  For applications with more than one Doppler signal
               last set of secured to tightening system use. “List other”   loss, this is the time of the last Doppler signal loss prior
               in the “Problems with or without arterial occlusion at   to hands off and “Done.”
               Done” receives the explanation.               6.  Hands off
          3.  Touch tightening system                          a.  Applier is done touching the tourniquet.
            a.  “Touch tightening system” time is when applier touches   b.  If the applier takes hands off but then realizes the Dop-
               tightening system with purpose to use the tightening sys-  pler signal isn’t gone and does additional tightening be-
               tem, for example not hand resting on tightening system   fore saying “Done,” then “hands off” is not the first time
               while still wondering what to do with rest of strap.  but the last time with hands off the tourniquet.
            b.  If applier only puts the tourniquet on the limb once and
               goes from placing the tourniquet around the limb to   PMID: 38112184; DOI: 10.55460/RPO1-CB79
               securing the tourniquet to using the tightening system