Page 42 - JSOM Winter 2019
P. 42

TABLE 48  M2/RMT-T/TX2/TX3 Usage                   TABLE 51  SAM-XT Time of Application
           Total                                              Total
           Score  Score           Usage             Citation  Score Score  Time (Speed) of Application  n =  Citation
                      2 uses in combat documented in medical            Application time – 70 seconds (± 30)
                  3   literature.                     37            2   to control bleeding.       80    40
                      No civilian usage documented in medical           Application time – 140 seconds
                  2   literature.                    n/a            3   with untrained firefighters to assess   12  52
           2.75                                                         mentored application.
                  3   Ranked 38.9% as preferred tourniquet   42
                      for arm application (n = 55).           3.60      Application time – 86 seconds
                      Ranked 30.2% as preferred tourniquet          4   median through 911 dispatch over-  31  54
                  3                                   42                the-phone mentoring.
                      for leg application (n = 55).
                                                                        Application time – >45 seconds for
                                                                    5   94% of participant.        32    53
          TABLE 49  M2/RMT-T/TX2/TX3 Logistics                          Application time – 9–15 seconds
           Total                                                    4   upper arm.                 10    56
           Score Score   Logistics
                                      6515-01-527-3841 (RMT-T)
                 5   NSN in DoD System.  6515-01-667-6027 (TX2)  It should be noted that two of the studies of the SAM-XT in-
                                       6515-01-667-6208 (TX3)  volved users who were either untrained on the device or were
                                          RMT-T $30.16       tele-mentored on how to apply the tourniquet. As such, these
           3.67  3   GSA Cost per Unit.    TX2 – $36.45
                                           TX3 – $37.95      studies resulted in extended application times. In contrast,
                                          RMT-T $35.95       for the two other studies involving the SAM-XT, the applica-
                     Commercial Cost per
                 3                         TX2 – $38.95      tions times for the SAMXT and CAT7 were not statistically
                     Unit.
                                           TX3 – $38.95      different.

          SAM Extremity Tourniquet (SAM-XT) –                TABLE 52  SAM-XT Simplicity of Application
          SCORE: 40.80.                                       Total  Part
                                                              Score Score Score    Ease of Use     n =  Citation
          The SAM-XT is a windlass (composite) tourniquet using Velcro
          adhesion to strap with single routed tension-locking buckle.   4   Mean ease-of-use score was 5   4  40
                                                                             (Easy) ± 0 out of 5.
                                                                   4.0       Ease of use reported >70%
          FIGURE 8  SAM Extremity Tourniquet.                            4   as easy with “click” of   12  52
                                                                             tourniquet.
                                                               9
                                                                   Part
                                                                   Score Score   Steps to Complete  n =  Citation
                                                                         5   5 steps.              n/a   57
                                                                   5.0       Mean of 2.9 ± 1 turns to
                                                                         5                         80    40
                                                                             occlusion.
                                                             TABLE 53  SAM-XT Pressures
                                                              Total
                                                              Score Score        Pressures         n =  Citation
                                                                    5   Mean of 186.07mmHg (± 62.957).  60  55
                                                               5    5   320mmHg (± 102).           80    40
          http://www.sammedical.com/products/sam-xt                 5   287mmHg (220–424).         31    54

          TABLE 50  SAM-XT Occlusion Efficacy                TABLE 54  SAM-XT Specifications
           Total                                              Total
           Score Score     Occlusion Efficacy   n =  Citation  Score Score       Specifications        Scoring
                 10  97.5% effectiveness.       80    40            1   Width – ≥1.5 inches   1.5 inches  1 – Yes /
                     100% occlusion with untrained                      (critical requirement)        –1 – No
                 9   firefighters to assess mentored   12  52           Length – ≥37.50 inches
                     application.                                   1   or provide 35 inches   35 inches   1 – Yes /
                     94% occlusion self-applied thigh                   circumferential (critical   circumferential  –1 – No
                 8   application by Law Enforcement   32  53            requirement).
           8.20      Officers.                                 5    1   Locking Mechanism/    Yes     1 – Yes /
                     100% occlusion medical-trained                     Method.                        0 – No
                 10  individuals using 911-dispatch   31  54        1   Time Recording.       Yes     1 – Yes /
                     mentoring.                                                                        0 – No
                     73% occlusion on HapMed Leg                        Weight – <8 ounces            1 – Yes /
                 4   Tourniquet Trainers simulators by   60  55     1   (critical for JFAK   3.8 oz   –1 – No
                     IDF Cadets.                                        inclusion).





          40  |  JSOM   Volume 19, Edition 4 / Winter 2019
   37   38   39   40   41   42   43   44   45   46   47