SYNCHRONIZED CARDIOVERSION
                                 Clinical Indications
         •  Unstable patient with tachycardia, dysrhythmia noted on monitor/EKG
         •  Patient who has failed conservative and/or chemical cardioversion
         •  Pulse present
                                  Contraindications
         •  None
                                     Procedure
         1. Ensure the patient is attached to a monitor/defibrillator with synchronized cardioversion
           capability.
         2. Time-permitting, ensure adequate IV/IO access is present. Ensure that unsynchronized
           cardioversion/defibrillation capabilities are present in case the patient degenerates into another
           dysrhythmia.
         3. Consider use of sedating medication (e.g., Midazolam 0.5–2mg q5min) prior to delivery
           of shock.
           Note: This step is not mandatory and should not delay the appropriate management of an
           emergent condition.
         4. Set energy level IAW appropriate ALS/PALS Protocol
         5. Select the Synchronized Cardioversion option. This should result in the machine displaying
           “SYNC” as well as tracking electrical activity (arrow or highlighted segment of EKG)
         6. Ensure all personnel are clear of the patient and that pilots are aware of cardioversion
         7. Press and hold the “Shock” button until energy is delivered (this may take several seconds for
           the machine to synchronize with the cardiac cycle. Shock is not immediately delivered as in
           defibrillation)
         8. If rhythm converts, monitor and treat as appropriate
         9. If it fails to convert, repeat steps 4–7 above using escalating energy levels. If the patient
           degenerates, treat as per appropriate ALS/PALS protocol.
           Note: Most machines require pushing the “SYNC” after each shock if synchronized
           cardioversion is to be repeated; failure to do so will result in delivery of an
           unsynchronized shock.
                        **Document procedure, results, and vital signs.**